A magnetic-bottle multi-electron-ion coincidence spectrometer

Akitaka Matsuda, Mizuho Fushitani, Chien Ming Tseng, Yasumasa Hikosaka, John H.D. Eland, Akiyoshi Hishikawa*

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

23 被引用数 (Scopus)

抄録

A novel multi-electron-ion coincidence spectrometer developed on the basis of a 1.5 m-long magnetic-bottle electron spectrometer is presented. Electrons are guided by an inhomogeneous magnetic field to a detector at the end of the flight tube, while a set of optics is used to extract counterpart ions to the same detector, by a pulsed inhomogeneous electric field. This setup allows ion detection with high mass resolution, without impairing the high collection efficiency for electrons. The performance of the coincidence spectrometer was tested with double ionization of carbon disulfide, CS2 → CS 22+e-e-, in ultrashort intense laser fields (2.8 × 1013W/cm2, 280 fs, 1030 nm) to clarify the electron correlation below the rescattering threshold.

本文言語英語
論文番号103105
ジャーナルReview of Scientific Instruments
82
10
DOI
出版ステータス出版済み - 2011/10

ASJC Scopus 主題領域

  • 器械工学

フィンガープリント

「A magnetic-bottle multi-electron-ion coincidence spectrometer」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。