Validation of differences in backscatter coefficients among four ultrasound scanners with different beamforming methods

Masaaki Omura*, Hideyuki Hasegawa, Ryo Nagaoka, Kenji Yoshida, Tadashi Yamaguchi

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

11 被引用数 (Scopus)

抄録

Purpose: The backscatter coefficient (BSC) indicates the absolute scatterer property of a material, independently of clinicians and system settings. Our study verified that the BSC differed among the scanners, transducers, and beamforming methods used for quantitative ultrasound analyses of biological tissues. Methods: Measurements were performed on four tissue-mimicking homogeneous phantoms containing spherical scatterers with mean diameters of 20 and 30 µm prepared at concentrations of 0.5 and 2.0 wt%, respectively. The BSCs in the different systems were compared using ultrasound scanners with two single-element transducers and five linear high- or low-frequency probes. The beamforming methods were line-by-line formation using focused imaging (FI) and parallel beam formation using plane wave imaging (PWI). The BSC of each system was calculated by the reference phantom method. The mean deviation from the theoretical BSC computed by the Faran model was analyzed as the benchmark validation of the calculated BSC. Results: The BSCs calculated in systems with different properties and beamforming methods well concurred with the theoretical BSC. The mean deviation was below ± 2.8 dB on average, and within the approximate standard deviation (± 2.2 dB at most) in all cases. These variations agreed with a previous study in which the largest error among four different scanners with FI beamforming was 3.5 dB. Conclusion: The BSC in PWI was equivalent to those in the other systems and to those of FI beamforming. This result indicates the possibility of ultra-high frame-rate BSC analysis using PWI.

本文言語英語
ページ(範囲)35-46
ページ数12
ジャーナルJournal of Medical Ultrasonics
47
1
DOI
出版ステータス出版済み - 2020/01/01

ASJC Scopus 主題領域

  • 放射線学、核医学およびイメージング

フィンガープリント

「Validation of differences in backscatter coefficients among four ultrasound scanners with different beamforming methods」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル