TY - JOUR
T1 - Surveillance evaluation of the standardization of assay values for serum total 25-hydroxyvitamin D concentration in Japan
AU - Ihara, Hiroshi
AU - Kiuchi, Sachiko
AU - Ishige, Takayuki
AU - Nishimura, Motoi
AU - Matsushita, Kazuyuki
AU - Satoh, Mamoru
AU - Nomura, Fumio
AU - Yamashita, Mine
AU - Kitajima, Isao
AU - Tsugawa, Naoko
AU - Okano, Toshio
AU - Hirota, Kouichi
AU - Miura, Masakazu
AU - Totani, Masayuki
AU - Hashizume, Naotaka
N1 - Publisher Copyright:
© The Author(s) 2018.
PY - 2018/11/1
Y1 - 2018/11/1
N2 - Background: To assess the vitamin D nutritional status, serum total 25-hydroxyvitamin D (25(OH)D) concentration is measured. We used six automated 25(OH)D immunoassays (AIAs) available in Japan and certified by the Vitamin D Standardization Program (VDSP) at the U.S. Center for Disease Control and Prevention to assess the concordance of the assay results. Methods: Serum total 25(OH)D concentrations in SRM 972a and 20 serum samples from patients were determined using three liquid chromatography-tandem mass spectrometry (LC-MS/MS) and six AIAs (pilot study), and an additional 110 serum samples were assessed by the six AIAs (surveillance study). The assay bias from the results of LC-MS/MS by Chiba University or consensus values (i.e. average of six AIAs) was estimated using the procedure described in CLSI document EP09-A3. Results: LC-MS/MS at Chiba University could completely separate 25(OH)D2, 25(OH)D3 and 3-epi-25(OH)D3, and the observed values including total 25(OH)D in SRM 972a were all within ±1·SD of the assigned values. All AIAs produced results greater than ±3·SD. In the pilot study, four of the six AIAs had an average percentage bias, as estimated by confidence interval (CI), larger than ±5% (acceptance criterion in CLSI); the bias converged from −6.5% to 3.2% after adjustment by LC-MS/MS. In the surveillance study, 25(OH)D concentrations in AIAs all adjusted to LC-MS/MS converged within ±5% from consensus values. However, some AIAs showed negative or positive bias from the consensus values. Conclusions: Current AIAs in Japan continue to lack standardization. Manufacturers should implement quality assurance strategies so that their values more closely align to those of standard reference material 972a.
AB - Background: To assess the vitamin D nutritional status, serum total 25-hydroxyvitamin D (25(OH)D) concentration is measured. We used six automated 25(OH)D immunoassays (AIAs) available in Japan and certified by the Vitamin D Standardization Program (VDSP) at the U.S. Center for Disease Control and Prevention to assess the concordance of the assay results. Methods: Serum total 25(OH)D concentrations in SRM 972a and 20 serum samples from patients were determined using three liquid chromatography-tandem mass spectrometry (LC-MS/MS) and six AIAs (pilot study), and an additional 110 serum samples were assessed by the six AIAs (surveillance study). The assay bias from the results of LC-MS/MS by Chiba University or consensus values (i.e. average of six AIAs) was estimated using the procedure described in CLSI document EP09-A3. Results: LC-MS/MS at Chiba University could completely separate 25(OH)D2, 25(OH)D3 and 3-epi-25(OH)D3, and the observed values including total 25(OH)D in SRM 972a were all within ±1·SD of the assigned values. All AIAs produced results greater than ±3·SD. In the pilot study, four of the six AIAs had an average percentage bias, as estimated by confidence interval (CI), larger than ±5% (acceptance criterion in CLSI); the bias converged from −6.5% to 3.2% after adjustment by LC-MS/MS. In the surveillance study, 25(OH)D concentrations in AIAs all adjusted to LC-MS/MS converged within ±5% from consensus values. However, some AIAs showed negative or positive bias from the consensus values. Conclusions: Current AIAs in Japan continue to lack standardization. Manufacturers should implement quality assurance strategies so that their values more closely align to those of standard reference material 972a.
KW - 25-Hydroxyvitamin D
KW - automated immunoassay
KW - liquid chromatography-tandem mass spectrometry
KW - standard reference material 972a
KW - standardization
UR - http://www.scopus.com/inward/record.url?scp=85045089794&partnerID=8YFLogxK
U2 - 10.1177/0004563218765570
DO - 10.1177/0004563218765570
M3 - 学術論文
C2 - 29534611
AN - SCOPUS:85045089794
SN - 0004-5632
VL - 55
SP - 647
EP - 656
JO - Annals of Clinical Biochemistry
JF - Annals of Clinical Biochemistry
IS - 6
ER -