Rapid determination of total mercury in treated waste water by cold vapor atomic absorption spectrometry in alkaline medium with sodium hypochlorite solution

Shigehiro Kagaya*, Yoshiharu Kuroda, Yuka Serikawa, Kiyoshi Hasegawa

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

15 被引用数 (Scopus)

抄録

Addition of a sodium hypochlorite solution (9.2% (w/v)) was effective to reduce a sulfide interference in determination of organic mercury, including methylmercury and phenylmercury, as well as a previously reported determination of inorganic mercury by cold vapor atomic absorption spectrometry (CVAAS) in an alkaline medium. Total mercury ranging from 0.17 to 33 μg L-1 in 15 mL of sample solutions containing up to 200 mg L-1 of sulfide can be determined without any serious interference by sulfide when 1 mL of the sodium hypochlorite solution was added after dilution of the sample solution to 25 mL. The proposed method was simple and rapid because no digestion processes were required for the determination of total mercury; the time required for the determination was only about 5 min. The proposed method was applicable to the analysis of treated waste water.

本文言語英語
ページ(範囲)554-557
ページ数4
ジャーナルTalanta
64
2
DOI
出版ステータス出版済み - 2004/10/08

ASJC Scopus 主題領域

  • 分析化学

フィンガープリント

「Rapid determination of total mercury in treated waste water by cold vapor atomic absorption spectrometry in alkaline medium with sodium hypochlorite solution」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル