抄録
This paper reports a parallel monitoring method for single cell respiration activity by using a microarrayed oxygen sensor. Optical oxygen sensor film was prepared by coating the mixture of 0.5g/l dichlorotris (1,10-phenanthroline) ruthenium (II) and 2.5% Nafion onto a diamond like carbon sputtered slide glass. Micro well (10 μm i.d.) array was prepared with carbon black doped epoxy resin onto the sensor film. The prepared microarrayed oxygen sensor enabled independent oxygen sensing in each micro well, and the difference of respiration activity of single cell (mouse lymphocyte) could be detected in a microarrayed oxygen sensor.
本文言語 | 英語 |
---|---|
ページ | 438-440 |
ページ数 | 3 |
出版ステータス | 出版済み - 2008 |
イベント | 12th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2008 - San Diego, CA, 米国 継続期間: 2008/10/12 → 2008/10/16 |
学会
学会 | 12th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2008 |
---|---|
国/地域 | 米国 |
City | San Diego, CA |
Period | 2008/10/12 → 2008/10/16 |
ASJC Scopus 主題領域
- 化学工学(その他)
- バイオエンジニアリング