Optimization of condition of ultrasonic beam for measurement of small change in thickness of arterial wall

Masaru Watanabe*, Hideyuki Hasegawa, Hiroshi Kanai

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

5 被引用数 (Scopus)

抄録

We previously developed a method for measuring small changes in thickness of the arterial wall during one cardiac cycle. Knowledge of this change in thickness is useful for in vivo assessment of the regional elasticity of the arterial wall. In this study, from computer simulations, it is found that measurement error depends on the distance of the ultrasonic beam from the center of the artery and it can be reduced by optimally setting the focal position. In basic experiments using a silicone rubber tube and in vivo experiments with a human carotid artery, it is found that by optimizing the focal position, measurement of the change in thickness becomes more robust against mispositioning of the ultrasonic beam. From these results, it is demonstrated that optimum focal positioning provides more robustness in measurement, even if there is arterial wall motion causing the position of the ultrasonic beam to deviate from the center of the artery.

本文言語英語
ページ(範囲)3613-3618
ページ数6
ジャーナルJapanese Journal of Applied Physics
41
5 B
DOI
出版ステータス出版済み - 2002/05

ASJC Scopus 主題領域

  • 工学一般
  • 物理学および天文学一般

フィンガープリント

「Optimization of condition of ultrasonic beam for measurement of small change in thickness of arterial wall」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル