Minute roughness measurement using phase tracking for arterial wall diagnosis non-invasively in vivo

Magnus Cinthio*, Hideyuki Hasegawa, Hiroshi Kanai

*この論文の責任著者

研究成果: 書籍の章/レポート/会議録会議への寄与査読

10 被引用数 (Scopus)

抄録

Early diagnosis of atherosclerosis is essential as many of the risk factors are life-style dependent. We suggest a new method to measure minute roughness of the size of micrometers of the arterial wall. The method was evaluated using three silicone phantoms sized 13 μm, 23 μm and 33 μm, respectively. The mean of the measured heights of the phantoms were 8.1 μm (SD 0.0), 23.3 μm (SD 0.2) and 29.6 μm (SD 0.1) in the forward direction, and 7.7 μm (SD 0.0), 21.9 μm (SD 0.2) and 27.3 μm (SD 3.6) in the backward direction, respectively. The phantom study shows very promising result and encourages to further evaluation and in vivo investigations.