Micra trans-catheter leadless pacemaker implantation in a patient with large right heart

Teruhiko Imamura*, Takahisa Koi, Naoya Kataoka, Shuhei Tanaka, Hiroshi Ueno, Koichiro Kinugawa

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

3 被引用数 (Scopus)

抄録

The use of leadless trans-catheter pacemakers is increasing particularly in the elderly population. However, its indication for those with anatomical anomaly remains unknown. We had a 75-year-old woman with atrial standstill and ventricular escape. Micra leadless pacemaker (Medtronic, Inc, Minneapolis, MN, USA) failed to be deployed due to too enlarged right atrium accompanied by atrial septal deficiency, followed by successful implantation of transvenous pacemaker lead by using SelectSecure lead (Medtronic) with a C315 delivery catheter that enhanced back-up force toward the ventricular septum against significant tricuspid regurgitation. The Micra is a promising system, but we should understand its limitations as well as alternative systems particularly for such an anatomical anatomy. <Learning objective: A Micra leadless trans-catheter pacemaker is a promising system with less invasiveness, particularly for elderly patients, but we should understand its technical limitation and consider alternative systems if necessary, particularly for those with enlarged right heart.>

本文言語英語
ページ(範囲)136-139
ページ数4
ジャーナルJournal of Cardiology Cases
24
3
DOI
出版ステータス出版済み - 2021/09

ASJC Scopus 主題領域

  • 循環器および心血管医学

フィンガープリント

「Micra trans-catheter leadless pacemaker implantation in a patient with large right heart」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。