Investigation on improving performance of adaptive beamformer by statistical analysis of ultrasonic echoes

Takumi Akamatsu, Michiya Mozumi, Masaaki Omura, Ryo Nagaoka, Hideyuki Hasegawa*

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

4 被引用数 (Scopus)

抄録

Minimum variance (MV) beamformers have been introduced in medical ultrasound imaging to improve image quality. In most cases, the MV beamformers have been investigated in terms of resolution improvement. However, the contrast-to-noise ratio (CNR) is also a clinically important metrics and gathers attention recently. In this study, we examined the diagonal loading parameter σ in MV beamforming and determined its appropriate value by evaluating image quality evaluation metrics including CNR. In order to further improve the image quality, a method for determining the value of σ based on the difference in statistical properties of received ultrasonic echo signals was also investigated. The phantom experimental results showed that the proposed method achieved a better CNR than the conventional MV beamformer while keeping resolution significantly better than that in delay-and-sum beamforming.

本文言語英語
論文番号SG1040
ジャーナルJapanese Journal of Applied Physics
61
SG
DOI
出版ステータス出版済み - 2022/07/01

ASJC Scopus 主題領域

  • 工学一般
  • 物理学および天文学一般

フィンガープリント

「Investigation on improving performance of adaptive beamformer by statistical analysis of ultrasonic echoes」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル