抄録
The singular value decomposition (SVD) based clutter filter is commonly applied to beamformed signals for the visualization of echo signals from flowing blood cells. In this paper, the SVD-based clutter filter is applied to signals directly acquired from ultrasonic elements before beamforming to be compared with the conventional strategy by evaluating contrast and standard deviation (SD) in the filtered images. As a result, the contrast was improved from 10.7 ± 3.6 dB to 18.3 ± 4.6 dB, and the SD was slightly improved from 3.78 ± 0.69 dB to 3.07 ± 0.74 dB in the measurement of a right jugular vein.
本文言語 | 英語 |
---|---|
論文番号 | SG1011 |
ジャーナル | Japanese Journal of Applied Physics |
巻 | 61 |
号 | SG |
DOI | |
出版ステータス | 出版済み - 2022/07 |
ASJC Scopus 主題領域
- 工学一般
- 物理学および天文学一般