Comparison between a noncontact uroflowmeter and a medical uroflowmeter in terms of uroflowmetry variables

Yoshio Kanayama, Takumi Kamimura, Shunsuke Ikegami, Hiroshi Kitamura, Kazuki Nakajima*

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

1 被引用数 (Scopus)

抄録

The present studyaimed to validate a noncontact uroflowmeter developed byus. The noncontact uroflowmeter was compared to a medical uroflowmeter in terms of uroflowmetryvariables. We used the uroflowmeters simultaneouslyto assess pseudourine and urine of healthymales, and compared the urinaryflow pattern, voiding time, voided volume, and average and maximum urinaryflow rates. Both uroflowmeters yielded similar urinaryflow patterns. In the case of pseudourine, the two uroflowmeters differed significantlyonlyin voiding time, which was significantlyshorter with the noncontact uroflowmeter than with the medical uroflowmeter. However, with regard to urine, no significant differences were observed in the terms of uroflowmetryvariables between the uroflowmeters. From these results, the possibilityof using the noncontact uroflowmeter for medical use has increased.

本文言語英語
ページ(範囲)54-60
ページ数7
ジャーナルTransactions of Japanese Society for Medical and Biological Engineering
58
1
DOI
出版ステータス出版済み - 2020

ASJC Scopus 主題領域

  • 生体医工学

フィンガープリント

「Comparison between a noncontact uroflowmeter and a medical uroflowmeter in terms of uroflowmetry variables」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。