An operating microscope with higher magnification and higher resolution for cerebral aneurysm surgery: Preliminary experience-technical note

Nobuhisa Matsumura*, Takashi Shibata, Emiko Hori, Hironaga Kamiyama, Mariko Tani, Soushi Okamoto, Michiya Kubo, Yukio Horie, Shunro Endo, Satoshi Kuroda

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

8 被引用数 (Scopus)

抄録

We describe a higher magnifying power operating microscope system to improve one method of highquality microsurgical clipping for cerebral aneurysm in some cases. This higher magnification is achieved by a new lens design in the optical system, which makes the image of the object very clear at high magnifications (distinctiveness of 7 μm). This higher-resolution operating microscope system provides the surgeon with higher-magnified images (at the maximum of more than 30× magnifications as each working distance) in the operating field. The magnifications can be changed from low power (2.9×) to high power (62.0×) depending on the circumstances in a given procedure. We have used this operating microscope system on 11 patients with microsurgical clipping for cerebral aneurysms. Microsurgical treatment could be performed safely and precisely. All aneurysms were treated without any technical complications. We think that the use of this microscope would have potential benefits for microsurgical treatment for cerebral aneurysms because of better visualization.

本文言語英語
ページ(範囲)497-501
ページ数5
ジャーナルNeurologia Medico-Chirurgica
54
6
DOI
出版ステータス出版済み - 2014

ASJC Scopus 主題領域

  • 外科
  • 臨床神経学

フィンガープリント

「An operating microscope with higher magnification and higher resolution for cerebral aneurysm surgery: Preliminary experience-technical note」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル