抄録
For the diagnosis of the early stages of atherosclerosis, it is important to evaluate the local acoustic characteristics of the arterial wall. For this purpose, it is necessary to increase the spatial resolution in the axial direction to several millimeters, which corresponds to the size of the macular lesion on the surface of the wall. We have proposed a method for measuring mall velocity signals on the intima and adventitia of the arterial wall from the skin surface using pulsive ultrasonic waves. The mall change in thickness of the arterial wall is obtained by integrating the difference between the two velocity signals on the intima and adventitia. The elastic property of the arterial wall is noninvasively evaluated from the change in thickness and the arterial inner pressure. In this paper, we evaluate the accuracy of the proposed method for measuring the small displacement. Moreover, we applied this method to evaluate the elastic property of the arterial wall of 50 patients and 8 healthy subjects.
本文言語 | 英語 |
---|---|
ページ(範囲) | 3101-3105 |
ページ数 | 5 |
ジャーナル | Japanese Journal of Applied Physics |
巻 | 37 |
号 | 5 SUPPL. B |
DOI | |
出版ステータス | 出版済み - 1998/05 |
ASJC Scopus 主題領域
- 工学一般
- 物理学および天文学一般