A laser interferometer accelerometer for vibration sensitive cryogenic experiments

R. Bajpai*, T. Tomaru, K. Yamamoto, T. Ushiba, N. Kimura, T. Suzuki, T. Yamada, T. Honda

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

6 被引用数 (Scopus)

抄録

Monitoring motion originating from ultra low-temperature cooling systems like cryocoolers is important for vibration sensitive cryogenic experiments like KAGRA. Since no high sensitivity commercial cryogenic accelerometers are available, we developed a compact self-calibrating accelerometer with a Michelson interferometer readout for cryogenic use. Change in calibration factor and drop in interferometer output originating from temperature drop were the main concerns which were tackled. Sensitivity of 3.38×10-11 m/Hz at 1 Hz was achieved at 300 K. The accelerometer was tested inside the KAGRA cryostat; showed stable operation down to 12 K in 0.1-100 Hz band with only 1% visibility drop. Our accelerometer can be employed in low vibration cryogenic environment for a multitude of applications.