体重計による尿流測定のためのアルゴリズム開発

野呂 剛史, 金 主賢, 藤田 徹也, 中島 一樹*

*この論文の責任著者

研究成果: ジャーナルへの寄稿学術論文査読

抄録

Acquisition of daily urination information leads to accurate diagnosis and treatment of dysuria. Conventional uroflowmeters have some problems about installation cost, sanitation management and measurement condition. As a method to solve them, uroflowmetry by the bath scale was proposed. In this study, flow rate change was calculated from weight change and smoothed by moving average to remove effect of body sway. First, pseudo-urination was performed on the bath scale and an optimum number for smoothing was determined. Next, urination was performed on the bath scale, and it was measured simultaneously with a uroflowmeter. The accuracy of 5 measurements was examined with 7 subjects. As a result, the bell-shaped urinary flow patterns followed to the patterns of uroflowmeter. The estimated voided volume (VVest) and maximum flow rate (Qmax) distributed within ±20% of the reference.

寄稿の翻訳タイトルDevelopment of an Algorithm for Uroflowmetry Using a Bath Scale
本文言語日本
ページ(範囲)64-70
ページ数7
ジャーナルライフサポート
31
2
DOI
出版ステータス出版済み - 2019/06/28

引用スタイル