Quartz-crystal gelation detector for the determination of fibrinogen concentration

H. Muramatsu*, E. Tamiya, M. Suzuki, I. Karube

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

65 Scopus citations

Abstract

A quartz crystal resonator was used for gelation monitoring and applied to the determination of fibrinogen concentration. The system consists of a 9-MHz AT-cut quartz crystal with a well- type cell, oscillator, frequency counter and microcomputer. The gelation time of fibrinogen on mixing with thrombin solution was determined by computer processing. The effect of temperature was avoided by the addition of aluminium oxide particles to the thrombin solution. The gelation time obtained in this system showed good linearity with fibrinogen concentration over the range 50-500 mg d l-1.

Original languageEnglish
Pages (from-to)321-326
Number of pages6
JournalAnalytica Chimica Acta
Volume217
Issue numberC
DOIs
StatePublished - 1989

ASJC Scopus subject areas

  • Analytical Chemistry
  • Biochemistry
  • Environmental Chemistry
  • Spectroscopy

Fingerprint

Dive into the research topics of 'Quartz-crystal gelation detector for the determination of fibrinogen concentration'. Together they form a unique fingerprint.

Cite this