Potential pitfalls of a fiducial marker-matching technique in carbon-ion radiotherapy for lung cancer

Shintaro Shiba, Jun Ichi Saitoh*, Daisuke Irie, Katsuyuki Shirai, Takanori Abe, Yoshiki Kubota, Makoto Sakai, Ryosuke Okada, Tatsuya Ohno, Takashi Nakano

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

8 Scopus citations

Abstract

Background/Aim: To analyze the accuracy of patient positioning and dose distribution quality using a fiducial marker-matching technique in carbon-ion radiotherapy (C-ion RT) for stage I lung cancer. Patients and Methods: Thirteen patients with 26 fiducial markers and 104 radiation fields were examined. Change in the fiducial marker position coordinates from the gross tumor volume center (Δm), and change in the water-equivalent path length of the chest wall (ΔWEL) were measured in each radiation field. Criterion for an acceptable change in dose distribution was defined as the percentage of D95 (%D95) at gross tumor volume greater than 90% of the prescribed dose. Results: Thirteen radiation fields (12.5%) were classified as unacceptable. Δm and ΔWEL had significant negative correlations with %D95 and thus were significant factors related to unacceptable irradiation fields. Conclusion: Patient positioning using a fiducial marker-matching technique resulted in 12.5% of radiation fields demonstrating unacceptable degradation due to Δm and ΔWEL.

Original languageEnglish
Pages (from-to)5673-5680
Number of pages8
JournalAnticancer Research
Volume37
Issue number10
DOIs
StatePublished - 2017/10

Keywords

  • Carbon-ion radiotherapy
  • Fiducial marker matching
  • Lung cancer

ASJC Scopus subject areas

  • Oncology
  • Cancer Research

Fingerprint

Dive into the research topics of 'Potential pitfalls of a fiducial marker-matching technique in carbon-ion radiotherapy for lung cancer'. Together they form a unique fingerprint.

Cite this