TY - JOUR
T1 - Non-harmonic analysis applied to optical coherence tomography imaging
AU - Cao, Xu
AU - Uchida, Tetsuya
AU - Hirobayashi, Shigeki
AU - Chong, Changho
AU - Morosawa, Atsushi
AU - Totsuka, Koki
AU - Suzuki, Takuya
PY - 2012/2
Y1 - 2012/2
N2 - A new processing technique called non-harmonic analysis (NHA) is proposed for optical coherence tomography (OCT) imaging. Conventional Fourier-domain OCT employs the discrete Fourier transform (DFT), which depends on the window function and length. The axial resolution of the OCT image, calculated by using DFT, is inversely proportional to the full width at half maximum (FWHM) of the wavelength range. The FWHM of wavelength range is limited by the sweeping range of the source in swept-source OCT and it is limited by the number of CCD pixels in spectraldomain OCT. However, the NHA process does not have such constraints; NHA can resolve high frequencies irrespective of the window function and the frame length of the sampled data. In this study, the NHA process is described and it is applied to OCT imaging. It is compared with OCT images based on the DFT. To demonstrate the benefits of using NHA for OCT, we perform OCT imaging with NHA of an onion skin. The results reveal that NHA can achieve an image resolution equivalent that of a 100-nm sweep range using a significantly reduced wavelength range. They also reveal the potential of using this technique to achieve high-resolution imaging without using a broadband source. However, the long calculation times required for NHA must be addressed if it is to be used in clinical applications.
AB - A new processing technique called non-harmonic analysis (NHA) is proposed for optical coherence tomography (OCT) imaging. Conventional Fourier-domain OCT employs the discrete Fourier transform (DFT), which depends on the window function and length. The axial resolution of the OCT image, calculated by using DFT, is inversely proportional to the full width at half maximum (FWHM) of the wavelength range. The FWHM of wavelength range is limited by the sweeping range of the source in swept-source OCT and it is limited by the number of CCD pixels in spectraldomain OCT. However, the NHA process does not have such constraints; NHA can resolve high frequencies irrespective of the window function and the frame length of the sampled data. In this study, the NHA process is described and it is applied to OCT imaging. It is compared with OCT images based on the DFT. To demonstrate the benefits of using NHA for OCT, we perform OCT imaging with NHA of an onion skin. The results reveal that NHA can achieve an image resolution equivalent that of a 100-nm sweep range using a significantly reduced wavelength range. They also reveal the potential of using this technique to achieve high-resolution imaging without using a broadband source. However, the long calculation times required for NHA must be addressed if it is to be used in clinical applications.
UR - http://www.scopus.com/inward/record.url?scp=84863144483&partnerID=8YFLogxK
U2 - 10.1143/JJAP.51.022503
DO - 10.1143/JJAP.51.022503
M3 - 学術論文
AN - SCOPUS:84863144483
SN - 0021-4922
VL - 51
JO - Japanese Journal of Applied Physics
JF - Japanese Journal of Applied Physics
IS - 2 PART 1
M1 - 022503
ER -