In vivo evaluation of the national cardiovascular center electrohydraulic total artificial heart

Eisuke Tatsumi*, Toru Masuzawa, Makoto Nakamura, Yoshiyuki Taenaka, Takashi Nishimura, Seiko Endo, Bin Zhang, Yukihide Kakuta, Masako Nakata, Tomomichi Nakamura, Tomohiro Nishinaka, Hisateru Takano, Kinji Tsukahara, Katsuya Tsuchimoto

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

We have been developing an electrohydraulic total artificial heart system. The system comprises an intrathoracic pumping unit composed of diaphragm type ellipsoidal blood pumps and an energy converter in addition to an electronics unit. The in vivo performance of the pumping unit was evaluated in a series of animal implantations with 3 calves weighing 62-85 kg. An interatrial shunt 4.5 mm in diameter was made in the atrial septum to compensate left-right imbalance. Two calves died early postoperatively, one of external controller power failure and the other of interatrial shunt stenosis due to thrombus formation. One calf, however, survived over 10 days under stable circulatory conditions. No abnormality was found in the oxygen metabolic condition or in major organ functions. The generation and dissipation of heat from the device was acceptable. This animal died of device malfunction caused by energy converter bearing breakdown. The device demonstrated a good anatomic fit without compromising the great vessels and adjacent tissues. It is concluded that the pumping unit has a sufficient in vivo basic performance although appropriate countermeasures are to be implemented against the detected problems concerning mechanical durability and interatrial shunt patency.

Original languageEnglish
Pages (from-to)242-248
Number of pages7
JournalArtificial Organs
Volume23
Issue number3
DOIs
StatePublished - 1999

Keywords

  • Anatomic fit
  • Blood pump
  • Electrohydraulic total artificial heart
  • Energy converter
  • Left-right balance
  • Total artificial heart

ASJC Scopus subject areas

  • Bioengineering
  • Medicine (miscellaneous)
  • Biomaterials
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'In vivo evaluation of the national cardiovascular center electrohydraulic total artificial heart'. Together they form a unique fingerprint.